Kippzonen CM 4 Pyranometer Manual do Utilizador Página 44

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 53
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 43
43
APPENDIX I PYRANOMETER CLASSIFICATION
ACCORDING TO WMO GUIDE 1996
Characteristics
High
quality
Good
quality
Moderate
quality
ISO 9060 classification
Secondary
standard
First class Second class
Response time (95 percent response) < 15 s < 30 s < 60 s
Zero offset:
(a) Response to 200 W/m
2
net thermal radiation
(ventilated)
(b) Response 5 K/h change
in ambient temperature
± 7 W/m
2
± 2 W/m
2
± 15 W/m
2
± 4 W/m
2
± 30 W/m
2
± 8 W/m
2
Resolution (smallest detectable
change)
± 1 W/m
2
± 5 W/m
2
± 10 W/m
2
Stability (change per year,
percentage of full scale)
± 0.8 ± 1.5 ± 3.0
Directional response of beam
radiation
(The range of errors caused by
assuming that the normal incidence
responsivity is valid for all directions
when measuring, from any direction,
a beam radiation whose normal
incidence irradiance is 1000 W/m
2
)
± 10 W/m
2
± 20 W/m
2
± 30 W/m
2
Temperature response (percentage
of maximum due to any change of
ambient temperature within an
interval of 50 K)
± 2 ± 4 ± 8
Non-linearity (percentage deviation
from the responsivity at 500 W/m
2
due to any change of irradiance
within the range 100 to 1000 W/m
2
)
± 0.5 ± 1 ± 3
Spectral sensitivity (percentage of
deviation of the product of spectral
absorptance and spectral
transmittance from the corresponding
mean within the range of 0.3 to 3 μm)
± 2 ± 5 ± 10
Tilt response (percentage deviation
from the responsivity at 0° tilt,
horizontal, due to change in tilt from
0° to 90° at 1000 W/m
2
irradiance)
± 0.5 ± 2 ± 5