CFR manual
9
Step 5 Connect the red and black signal measurement cables to either the back or front input terminals of
the voltmeter and to the signal output terminals near the lamp pillar.
figure 5 Signal cables between voltmeter and signal output terminals
Step 6 The installation will be complete when the mains power cord is connected to an AC power outlet.
The lamp of the 230 VAC / 50 Hz Calibration Facility the lamp can be turned on and off using the switch on
the power strip.
Finally the correct lamp height should be determined by measuring the irradiance with a reference
pyranometer at the left or right position of the turntable. Adjust the lamp height such that the irradiance
should be in the range 500 600 W/m
2
.
figure 6 Complete calibration facility