Kippzonen CFR Calibration Facility Manual do Utilizador Página 19

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 29
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 18
CFR manual
18
3.4 Calibration Form
An example of a calibration form is given below.
Test pyranometer model:
CMP 6
Serial number:
060002
Reference pyranometer model:
CMP 6
Serial number:
060001
Sensitivity: 10.00 V/Wm
-2
(SR)
Calibration performed by:
J. Smith
Date: 20 September 2006 Signature
Reference pyranometer
Test pyranometer
Position 1
Illuminated
R 1980
T 2980
Darkened
RE 5
TE 10
R1=R-RE 1975
T1=T-TE 2970
Position 2
Illuminated
R 2100
T 2820
Darkened
RE 2
TE 8
R2=R-RE 2098
T1=T-TE 2812
Test pyranometer sensitivity
)21(
)21(
RR
TT
SRST
+
+
=
14.20 V/Wm
-2
Lamp stability
0.98 <
)22(
)11(
TR
TR
< 1.02 0.994 OK/Rejected
Vista de página 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 28 29

Comentários a estes Manuais

Sem comentários