Kippzonen Brewer MkIII Manual do Utilizador Página 86

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 136
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 85
MKIII OPERATOR’S MANUAL
83
DCFJJJ.nnn - Dispersion constants
The Dispersion Constants are used to calculate the ozone wavelength of the exit slits.
#
Example
Name
1
Disp
Header
2
2856.96
intercept for slit 1
3
7.674577E-02
slope for slit 1
4
-7.251786E-07
quadratic for slit 1
5
2896.561
intercept for slit 2
6
7.600413E-02
slope for slit 2
7
-7.387072E-07
quadratic for slit 2
8
2933.527
intercept for slit 3
9
.0751006
slope for slit 3
10
-7.337653E-07
quadratic for slit 3
11
2968.578
intercept for slit 4
12
7.440717E-02
slope for slit 4
13
-7.512483E-07
quadratic for slit 4
14
3003.31
intercept for slit 5
15
7.325987E-02
slope for slit 5
16
-7.065609E-07
quadratic for slit 5
17
2823.907
intercept for mercury exit slit
18
.0774763
slope for mercury exit slit
19
-7.259538E-07
quadratic for mercury exit slit
20-36
-
Not used
UMKSETUP.nnn - Umkehr setup file
#
Example
1
SXUNI4M.DAT
2
STDTABS.DAT
3
O3TABLE.DAT
4
CQMS.DAT
5
4.1099 2.3155 1.5600 0.8637
6
0.6760 0.3187 0.1490 0.0776 .. ozone coefficients for #039 STN TMO
7
3 6 11 12 1 1.0 812.0
8
999 1 34.838 0
9
0 ... debug file
1 - 4. Tables used in the calculation of Umkehrs.
5 - 6. Ozone coefficients
7. This line has 7 parameters:
1 Minimum starting zenith angle
2 Maximum starting zenith angle
3 Minimum ending zenith angle
4 Maximum ending zenith angle
5 Output control flag
6 Ozone scaling factor (used with Dobson data)
7 Surface pressure of the station (in mmHg)
8. This line has 4 parameters:
1 Station identifying number (3 digits)
2 Parameter no longer used (set to 1)
3 Latitude of station - used for first guess
4 Output control flag (set to 0)
9. Flag for debugging purposes only.
APPENDIX B CONFIGURATION FILES
Vista de página 85